Article No
11141680
Baumer
Pressure sensors
CombiPress PFMN-54.BC3R.A1551.41212.4500 Pressure sensor
Baumer CombiPress PFMN-54.BC3R.A1551.41212.4500 features low (up to 10 bar) pressure range, G 1/2 A DIN 3852-E connection variants and AISI 316L (1.4435) electropolished wetted parts material, membrane.
Technical Specifications
Article No
11141680
Product Description
- Programmable by touch screen
- Cavity-free process connection
- Resistant to all conventional CIP cleaning media
- Compact installation from G 1/2 A
- Optionally available with relay outputs
- Absolute pressure, relative pressure and vacuum measurement
- Easy and full programmable with FlexProgrammer 9701
Detailed Technical Data
Process connection
Connection variants
G 1/2 A DIN 3852-E
Wetted parts material, membrane
AISI 316L (1.4435)
COMPLIANCE AND APPROVALS
Explosion protection
- ATEX II 1G Ex ia IIC T5
- ATEX II 1D Ex ia IIIC T100 °C Da
PERFORMANCE CHARACTERISTICS
Pressure type
Relative (gauged)
Power-up time
< 10 s
Sample time
≤ 0.3 s
Measuring range
-1 … 5 bar
Max. turndown ratio
10 : 1
Max. measurement error
- ± 0.25 % FSR , up to 4:1 turndown ratio
- Including zero-point and span error, nonlinearity (by terminal base line), hysteresis and non-repeatability (EN 61298-2)
- For turndown, multiply this value by the applied turndown ratio
Temperature coefficient
- ≤ 0.05 % FSR/10 K , measuring span
- ≤ 0.05 % FSR/10 K , zero point
Long term stability
≤ 0.1 % FSR/a , IEC 770 6.3.2
Rise time (%10 … %90)
≤ 0.3 s
Standard error of measurement (BFSL)
- ± 0.1 % FSR , up to 4:1 turndown ratio
- Including non-linearity, hysteresis and non-repeatability according BFSL
- For turndown, multiply this value by the applied turndown ratio
Compensated temperature range
-40 … 85 °C