Article No
11278520
Measurement Range
-1 … 5 bar
Initial Accuracy
≤ 0.25 - 0.3 % FS
Baumer
Pressure sensors
PP56H-2.BB77R.D214.442021.0000 Pressure sensor
Baumer PP56H-2.BB77R.D214.442021.0000 features low (up to 10 bar) pressure range, relative (gauge) pressure type and G 1 A hygienic process connection.
Technical Specifications
Article No
11278520
Measurement Range
-1 … 5 bar
Initial Accuracy
≤ 0.25 - 0.3 % FS
Process Connection
G 1 A hygienic
Product Description
- Resistant to all common CIP cleaning media
- Condensate-proof measuring cell
- IO-Link dual channel output (in parallel to 4 ... 20 mA)
- Space-saving installations
- Easy parameterization via display, IO-Link and FlexProgram
- Rotatable display with 28 steps over 315° for threaded pressure connections
Detailed Technical Data
IO-LINK INTERFACE
SIO-mode
Yes
Baud rate
38,4 kbaud (COM2)
Dual Channel
IO-Link / SIO Mode
Device profile
Smart Sensor Profile
Dual Channel 2
Analogue / SIO Mode
Cycle time
≥ 4 ms
IO-Link port type
Class A
IO-Link version
1.1
Process data length
48 bit
Process data (cyclic)
- Pressure
- Switch state
- State of alert
Diagnostic data (acyclic)
- Measuring cell temperature
- CPU temperature
- Operation
Adjustable data (acyclic)
- Offset settings
- Switch parameters
- Current output limit values
- Display configuration
- Tags
- Password for access
Process connection
Wetted parts material
AISI 316L (1.4404)
Connection variants
G 1 A hygienic
Wet part material, membrane
AISI 316L (1.4435)
Performance characteristics
Pressure type
Relative (gauged)
Measuring range
-1 … 5 bar
Max. turndown ratio
5 : 1
Max. measurement error
- ± 0.3 % FSR , within compensated temperature range
- Including zero-point and span error, nonlinearity (by terminal base line), hysteresis and non-repeatability (EN 61298-2)
- For turndown, multiply this value by the applied turndown ratio
Temperature coefficient
- ≤ 0.1 % FSR/10 K @ -40 ... -20 °C
- ≤ 0.1 % FSR/10 K @ > 70 °C
- "Operating conditions" see section
Power-on time
< 600 ms @ 24 V
Long-term stability
≤ 0.2 % FSR/a
Standard error of measurement (BFSL)
- ± 0.12 % FSR , within compensated temperature range
- Including non-linearity, hysteresis and non-repeatability according BFSL
- For turndown, multiply this value by the applied turndown ratio
Compensated temperature range
-20 … 70 °C
Documents
description
Data sheet
PDF · Baumer_PP56H-2.BB77R.D214.442021.0000_EN_20251006_DS.pdf
open_in_new Open
menu_book
Operating manual
PDF · Baumer_PP5x_IODD_Parameter_and_Process_Data_MA_EN_1808_WEB.pdf
open_in_new Open
verified
Type Examination Certificate / Type approval – MTTF
PDF · Baumer_MTTF_PP56H-x.xxxxx_.xxxx_.xxxxx1_.xxxx_.pdf
open_in_new Open
verified
Type Examination Certificate / Type approval
PDF · Baumer_PFMH_PP56H_EN_CoC_81334594.pdf
open_in_new Open
restaurant
Food Regulatory Declaration of Harmlessness
PDF · Baumer_PP56H_PFMx_ML_FRDoH_81222146.pdf
open_in_new Open
verified
Type Examination Certificate / Type approval – MTTF
PDF · Baumer_MTTF_PP56(H)-x.xxxxx.xxxx.xxxxx1.xxxx.pdf
open_in_new Open
verified
Type Examination Certificate / Type approval
PDF · Baumer_PT55+PP20+PP54+PP56_CoC_81458954.pdf
open_in_new Open
article
Technical Reports / White Paper
PDF · Baumer_EHEDG_Position_Paper_201906.pdf
open_in_new Open
article
Technical Reports / White Paper
PDF · Baumer_Analog_Sensor_Interfaces_Guideline_V1.00_EN.pdf
open_in_new Open
article
Technical Reports / White Paper
PDF · Baumer_Switching_Output_Guideline_V1.00_EN.pdf
open_in_new Open
article
Technical Reports / White Paper
PDF · Baumer_Definitions of Sensor Properties_V1.1_EN.pdf
open_in_new Open